68 | Dome CXtra
NOTE: Look for the nominal value of the ambient light. This is a
record of the most recent measurement and provides a comparison
value for the next medical physicist survey.
If the reading room configuration has changed, the source of the
problem must be identified and corrective action taken before any
further mammographic images are reviewed or interpreted using
the Dome display.
To record the viewing conditions
1 Click Viewing Condition Check and Setting button on
the test wizard.
2 Enter the room description and room trait.
3 Load the drawing of the valid reading room configuration.
4 Click Submit to finish.